Laboratory Licensed by ema® to perform NRF-PEMEX
corrosion and accelerated weathering tests.
Laboratory Validated by Volkswagen to perform tests under the Standard
TL 211 & TL 226.
Datacolor ELREPHO™
Quality color measurement for the paper industry and other industries requiring diffuse /0° measurements.
Overview
The Datacolor ELREPHO is a high performance spectrophotometer designed for the paper industry and suitable for either the reference laboratory or production color control.
Designed For:
Laboratory and Quality Control for Pulp, Paper and Paper components.
PRODUCT DATA SHEET
Feature - Datacolor ELREPHO™
Instrument Type - dual-beam spectrophotometer
Measurement Geometry - diffuse illumination and 0° viewing
Illumination Source - pulsed xenon filtered to approximate D65
Sphere Diameter - 152 mm / 6.0 in
Spectral Analyzer - proprietary SP2000 analyzer with dual 256 diode array and high resolution holographic grating
Wavelength Range - 360 nm to 700 nm
Reporting Interval - 10 nm
Effective Bandwidth - 10 nm
Wavelength Resolution - 2 nm
Photometric Range: - 0 to 200%
Photometric Resolution - 0.003%
Black Trap - high performance
20 Read Repeatability On The White Tile Using Dual Flash (CIELAB) - 0.02 (max)
Inter-instrument Agreement 1(CIELAB) - 0.4 (maximum)
0.2 (average)
Lens - 3 position auto zoom
XLAV Aperture plate - 34 mm illuminated and 30 mm measured
SAV Aperture plate - 9 mm illuminated and 5 mm measured
USAV Aperture plate - 6.5 mm illuminated and 2.5 mm measured
Automatic UV Control - automatic UV calibration for the measurement of fluorescent specimens with UV cutoff filters at 395 nm, 420 nm and 460 nm
Vertical Mount - includes peephole sample viewer and pedestal
sample holder
Height - 597 mm / 23.5 in
Width - 312 mm / 12.3 in
Depth - 381 mm / 15.0 in
Weight - 17.24 kg / 38 lb
Power Requirements - 85 to 264 VAC, 47 to 63 Hz, 80 VA peak, 35 VA typical
Absolute Operating Range - 5° to 40° C, 5% to 85% non-condensing relative humidity
Interface - RS-232 9600/19200 baud